Skip to content

Measurement Problems

This section addresses issues observed in the measured size distributions and concentrations.

No Cluster Ions Visible

Symptom: The data does not show the presence of cluster ions (i.e., small ions).

Particle Loss Before Instrument Entry

Cause: Particles are lost in the inlet tubing. This can be due to an improperly grounded inlet or an excessively long inlet tube.

Solution: Ensure the instrument inlet is properly grounded and use the shortest possible inlet tube.

Particle Loss Inside Instrument

Cause: Particles are lost internally before reaching the mobility analyzer. This is often due to incorrect electrical connections in the sample preconditioning unit.

Solution:

  • Verify that all plugs in the sample preconditioning unit are connected to the correct sockets.
  • Confirm a solid electrical connection between the filter electrode and its corresponding socket.
  • Check the electrical connections of other components (charger needle, charger current electrode, post-filter electrode, discharger needle, discharger current electrode). Note that most of these faults would also trigger other diagnostic warnings.

Incorrect Offset Measurement

Cause: The "offset" operating mode is not functioning correctly, leading to an incorrect offset value being subtracted from the "ions" or "particles" mode signal.

Solution: Investigate and resolve any issues with the offset measurement mode.

Missing 9V Central Electrode Voltage

Cause: The 9V supply to the central electrode may be missing.

Solution: Check the relevant cable connections. If the issue persists, contact Airel Support.

Significant Difference Between Positive and Negative Concentrations

Symptom: There is a large, unexpected difference between the measured positive and negative particle/ion concentrations.

Solution:

  1. Check the software for any diagnostic warnings.
  2. Verify that all airflow tubes are correctly and securely connected, especially the sheath flow tubes running from the sheath air filters to the top of the analyzer.